ASTM F543-07e1

$32.50

Format: PDF      In Stock

 

Sale!
-50%

ASTM F543-07e1

$32.50

Standard Specification and Test Methods for Metallic Medical Bone Screws
standard by ASTM International, 06/01/2007

Format: PDF      In Stock

 

Category:

NOTE: Our website provide PDF immediately download, sometimes when the payment is processed and you did not receive the file, please feel free to connect with us.

Description

1.1 This specification provides requirements for materials, finish and marking, care and handling, and the acceptable dimensions and tolerances for metallic bone screws that are implanted into bone. The dimensions and tolerances in this specification are applicable only to metallic bone screws described in this specification.

1.2 This specification provides performance considerations and standard test methods for measuring mechanical properties in torsion of metallic bone screws that are implanted into bone. These test methods may also be applicable to other screws besides those whose dimensions and tolerances are specified here. The following annexes are included:

1.2.1 Annex A1Test Method for Determining the Torsional Properties of Metallic Bone Screws.

1.2.2 Annex A2Test Method for Driving Torque of Medical Bone Screws.

1.2.3 Annex A3Test Method for Determining the Axial Pullout Strength of Medical Bone Screws.

1.2.4 Annex A4Test Method for Determining the Self-Tapping Performance of Self-Tapping Medical Bone Screws.

1.2.5 Annex A5Specifications for Type HA and Type HB Metallic Bone Screws.

1.2.6 Annex A6Specifications for Type HC and Type HD Metallic Bone Screws.

1.2.7 Annex A7Specifications for Metallic Bone Screw Drive Connections.

1.3 This specification is based, in part, upon ISO 5835, ISO 6475, and ISO 9268.

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.5 This standard may involve the use of hazardous materials, operations, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

A7.1.1 This annex specifies the requirements for a variety of drive connections that can be implemented in metallic medical bone screws. The screw’s drive connection supplies the interconnection that is typically used in orthopedic surgery for inserting and removing bone screws.

Product Details

Published:
06/01/2007
Number of Pages:
20
File Size:
1 file , 410 KB
Redline File Size:
2 files , 720 KB